User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Transmation 1040 Verification (1 year) w/Ronan X86 DATE: 29-Nov-94 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 50% NUMBER OF TESTS: 32 NUMBER OF LINES: 178 CONFIGURATION: Fluke 5700A STANDARD: X86 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # Source: Transmation 1040/1040SP Instruction Manual (I.S. No. 100724-905) # The 90 day specifications of the 5700A are used in TUR computations. 1.001 STD X86 1.002 ASK- R P F W 1.003 ASK+ D N K S U M C L 1.004 MMFC C2 1.005 HEAD EQUIPMENT SETUP 1.006 DISP [32] [27][91]1m[27][91]5mWARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Ambient temperature: 18C - 28C. 1.007 DISP Relative humidity: less than 80%. 1.007 DISP Allow the UUT to stabilize. 1.008 DISP Lift the toggle switch and raise it to test the battery. 1.009 EVAL Did the battery check out at greater than 4.7 Volts? 2.001 DISP [32] Connect the [27][91]5m5700A[27][91]0m and the UUT. 2.001 DISP [32][27][91]1m 5700A TO Transmation 1040 2.001 DISP [32] OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ V.IN/OUT + 2.001 DISP [32] OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ V.IN/OUT - 2.001 DISP [32] AUX CURRENT OUTPUT ÄÄÄÄÄÄ mA IN + 2.002 HEAD {DC VOLTAGE TEST} 2.003 DISP Turn on the Transmation. 2.004 DISP Select 0-11V IN on the Transmation. 2.005 HEAD DC VOLTAGE MEASUREMENT TEST: {0-11V Range} 2.006 5700 11 0.000V 0.03% 0.04/ 2W #! Test Tol 0.0044, Sys Tol 7.5e-007, TUR 5866.667 (>= 4.00). 3.001 5700 11 2.000V 0.03% 0.04/ 2W #! Test Tol 0.005, Sys Tol 1.32e-005, TUR 378.788 (>= 4.00). 4.001 5700 11 5.000V 0.03% 0.04/ 2W #! Test Tol 0.0059, Sys Tol 2.9e-005, TUR 203.448 (>= 4.00). 5.001 5700 11 7.000V 0.03% 0.04/ 2W #! Test Tol 0.0065, Sys Tol 3.9e-005, TUR 166.667 (>= 4.00). 6.001 5700 11 10.000V 0.03% 0.04/ 2W #! Test Tol 0.0074, Sys Tol 5.4e-005, TUR 137.037 (>= 4.00). 7.001 HEAD {DC MILLIVOLTAGE TEST} 7.002 DISP Select 0-110mV IN on the Transmation. 7.003 HEAD DC VOLTAGE MEASUREMENT TEST: {0-110mV Range} 7.004 5700 110 0.00mV 0.06% 0.06/ 2W #! Test Tol 6.6e-005, Sys Tol 7.5e-007, TUR 88.000 (>= 4.00). 8.001 5700 110 20.00mV 0.06% 0.06/ 2W #! Test Tol 7.8e-005, Sys Tol 8.9e-007, TUR 87.640 (>= 4.00). 9.001 5700 110 50.00mV 0.06% 0.06/ 2W #! Test Tol 9.6e-005, Sys Tol 1.1e-006, TUR 87.273 (>= 4.00). 10.001 5700 110 70.00mV 0.06% 0.06/ 2W #! Test Tol 0.000108, Sys Tol 1.24e-006, TUR 87.097 (>= 4.00). 11.001 5700 110 100.00mV 0.06% 0.06/ 2W #! Test Tol 0.000126, Sys Tol 1.45e-006, TUR 86.897 (>= 4.00). 12.001 HEAD {DC CURRENT TESTS} 12.002 DISP Select mA IN on the Transmation 12.003 HEAD DC CURRENT TESTS: {0-50mA Range} 12.004 5700 50 0.00mA 0.06% 0.12/ 2W #! Test Tol 6e-005, Sys Tol 1e-008, TUR 6000.000 (>= 4.00). 13.001 5700 50 4.00mA 0.06% 0.12/ 2W #! Test Tol 6.24e-005, Sys Tol 3e-007, TUR 208.000 (>= 4.00). 14.001 5700 50 12.00mA 0.06% 0.12/ 2W #! Test Tol 6.72e-005, Sys Tol 7e-007, TUR 96.000 (>= 4.00). 15.001 5700 50 20.00mA 0.06% 0.12/ 2W #! Test Tol 7.2e-005, Sys Tol 1.1e-006, TUR 65.455 (>= 4.00). 16.001 5700 50 50.00mA 0.06% 0.06/ 2W #! Test Tol 6e-005, Sys Tol 4e-006, TUR 15.000 (>= 4.00). 17.001 HEAD {DC VOLTAGE OUTPUT TESTS} 17.002 DISP [32] Connect the [27][91]5mRONAN[27][91]0m and the UUT. 17.002 DISP [32][27][91]1m RONAN TO Transmation 1040 17.002 DISP [32] INPUT V + ÄÄÄÄÄÄÄÄÄ V.IN/OUT + 17.002 DISP [32] INPUT V - ÄÄÄÄÄÄÄÄÄ V.IN/OUT - 17.003 DISP Turn on the Ronan. 17.004 DISP Select the 10V range on the Ronan. 17.005 DISP Select 0 - 11V OUT on the Transmation. 17.006 HEAD DC VOLTAGE OUTPUT TEST: {0-11V Range} 17.007 DISP Dial in zero volts on the Transmation. 17.008 MEMI Enter the reading from the lefthand Ronan display. 17.009 ACC 10 0.000V 0.02% 0.01/ 17.010 MEMC 11 V 0.03% 0.04/ #! Test Tol 0.0044, Sys Tol 0.001, TUR 4.400 (>= 4.00). 18.001 DISP Dial in 2 volts on the Transmation. 18.002 MEMI Enter the reading from the lefthand Ronan display. 18.003 ACC 10 2.000V 0.02% 0.01/ 18.004 MEMC 11 V 0.03% 0.04/ #! WARNING: Test Tol 0.005, Sys Tol 0.0014, TUR 3.571 (< 4.00). 19.001 DISP Dial in 5 volts on the Transmation. 19.002 MEMI Enter the reading from the lefthand Ronan display. 19.003 ACC 10 5.000V 0.02% 0.01/ 19.004 MEMC 11 V 0.03% 0.04/ #! WARNING: Test Tol 0.0059, Sys Tol 0.002, TUR 2.950 (< 4.00). 20.001 DISP Dial in 7 volts on the Transmation. 20.002 MEMI Enter the reading from the lefthand Ronan display. 20.003 ACC 10 7.000V 0.02% 0.01/ 20.004 MEMC 11 V 0.03% 0.04/ #! WARNING: Test Tol 0.0065, Sys Tol 0.0024, TUR 2.708 (< 4.00). 21.001 DISP Dial in 10 volts on the Transmation. 21.002 MEMI Enter the reading from the lefthand Ronan display. 21.003 ACC 10 10.000V 0.02% 0.01/ 21.004 MEMC 11 V 0.03% 0.04/ #! WARNING: Test Tol 0.0074, Sys Tol 0.003, TUR 2.467 (< 4.00). 22.001 DISP Select the 100mV range on the Ronan. 22.002 DISP Select 0 - 110mV OUT on the Transmation. 22.003 HEAD DC VOLTAGE OUTPUT TEST: {0-110mV Range} 22.004 DISP Dial in zero volts on the Transmation. 22.005 MEMI Enter the reading from the lefthand Ronan display. 22.006 ACC 100 0.000mV 0.02% 0.01/ 22.007 MEMC 110 mV 0.06% 0.06/ #! Test Tol 0.066, Sys Tol 0.01, TUR 6.600 (>= 4.00). 23.001 DISP Dial in 20 millivolts on the Transmation. 23.002 MEMI Enter the reading from the lefthand Ronan display. 23.003 ACC 100 20.00mV 0.02% 0.01/ 23.004 MEMC 110 mV 0.06% 0.06/ #! Test Tol 0.078, Sys Tol 0.014, TUR 5.571 (>= 4.00). 24.001 DISP Dial in 50 millivolts on the Transmation. 24.002 MEMI Enter the reading from the lefthand Ronan display. 24.003 ACC 100 50.00mV 0.02% 0.01/ 24.004 MEMC 110 mV 0.06% 0.06/ #! Test Tol 0.096, Sys Tol 0.02, TUR 4.800 (>= 4.00). 25.001 DISP Dial in 70 millivolts on the Transmation. 25.002 MEMI Enter the reading from the lefthand Ronan display. 25.003 ACC 100 70.00mV 0.02% 0.01/ 25.004 MEMC 110 mV 0.06% 0.06/ #! Test Tol 0.108, Sys Tol 0.024, TUR 4.500 (>= 4.00). 26.001 DISP Dial in 100 millivolts on the Transmation. 26.002 MEMI Enter the reading from the lefthand Ronan display. 26.003 ACC 100 100.00mV 0.02% 0.01/ 26.004 MEMC 110 mV 0.06% 0.06/ #! Test Tol 0.126, Sys Tol 0.03, TUR 4.200 (>= 4.00). 27.001 HEAD {DC CURRENT OUTPUT TESTS} 27.002 DISP [32] Connect the [27][91]5mRONAN[27][91]0m and the UUT. 27.002 DISP [32][27][91]1m RONAN TO Transmation 1040 27.002 DISP [32] INPUT mA+ ÄÄÄÄÄÄÄÄÄ mA OUT+ 27.002 DISP [32] INPUT mA- ÄÄÄÄÄÄÄÄÄ mA OUT- 27.003 DISP Select the 100 mA range on the Ronan. 27.004 DISP Select mA OUT on the Transmation. 27.005 HEAD DC CURRENT OUTPUT TEST: {0-50mA Range} 27.006 DISP Dial in zero mA on the Transmation. 27.007 MEMI Enter the reading from the lefthand Ronan display. 27.008 ACC 100 0.000mA 0.03% 0.01/ 27.009 MEMC 50 mA 0.06% 0.12/ #! Test Tol 0.06, Sys Tol 0.01, TUR 6.000 (>= 4.00). 28.001 DISP Dial in 4 mA on the Transmation. 28.002 MEMI Enter the reading from the lefthand Ronan display. 28.003 ACC 100 4.000mA 0.03% 0.01/ 28.004 MEMC 50 mA 0.06% 0.12/ #! Test Tol 0.0624, Sys Tol 0.0112, TUR 5.571 (>= 4.00). 29.001 DISP Dial in 12 mA on the Transmation. 29.002 MEMI Enter the reading from the lefthand Ronan display. 29.003 ACC 100 12.000mA 0.03% 0.01/ 29.004 MEMC 50 mA 0.06% 0.12/ #! Test Tol 0.0672, Sys Tol 0.0136, TUR 4.941 (>= 4.00). 30.001 DISP Dial in 20 mA on the Transmation 30.002 MEMI Enter the reading from the lefthand Ronan display. 30.003 ACC 100 20.000mA 0.03% 0.01/ 30.004 MEMC 50 mA 0.06% 0.12/ #! Test Tol 0.072, Sys Tol 0.016, TUR 4.500 (>= 4.00). 31.001 DISP Dial in 50 mA on the Transmation. 31.002 MEMI Enter the reading from the lefthand Ronan display. 31.003 ACC 100 50.000mA 0.03% 0.01/ 31.004 MEMC 50 mA 0.06% 0.06/ #! WARNING: Test Tol 0.06, Sys Tol 0.025, TUR 2.400 (< 4.00). 32.001 DISP Turn off the Transmation. 32.002 DISP Turn off the Ronan. 32.003 HEAD 32.004 END #! T.U.R.s less than 4.00: 5 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.